Table of Contents
CHAPTER 1 — GENERAL INFORMATION
1-1. SCOPE
This manual covers the 4-wire Kelvin measurement method for precision resistance calibration. By separating current-forcing leads from voltage-sensing leads, lead and contact resistance is eliminated from the measurement. Applicable for resistances from 0.001 Ω to 10k Ω.
CHAPTER 2 — THEORY OF OPERATION
2-1. KELVIN PRINCIPLE
In a standard 2-wire resistance measurement, the DMM forces current through the same leads it uses to measure voltage. Lead resistance (typically 0.1–0.5 Ω per lead) adds directly to the measured value, introducing significant error for low resistances.
In 4-wire Kelvin measurement, dedicated current-force leads (F+, F−) carry the test current, and separate voltage-sense leads (S+, S−) measure the voltage drop across the unknown resistor only. The voltage-sense leads carry negligible current (DMM input impedance >10 MΩ), so their resistance does not affect the measurement.
CHAPTER 3 — MATERIALS AND CONSTRUCTION
3-1. KELVIN PROBES AND FIXTURES
| Item | Description |
|---|---|
| DMM with 4-wire mode | Fluke 15B+ does NOT have 4-wire; requires Fluke 87V or bench meter |
| Kelvin clip probes | Paired force/sense clips per terminal (4 total) |
| Reference resistors | Certified 0.01% standard resistors: 0.1, 1, 10, 100, 1k Ω |
| 4-wire socket | DUT mounting with separated F+, F−, S+, S− connections |
CHAPTER 4 — ASSEMBLY PROCEDURES
4-1. DIY KELVIN FIXTURE
- Build a simple current source: LM334 or constant-current diode set to 1 mA through unknown resistor.
- Measure voltage across resistor with separate high-impedance voltmeter leads.
- Rx = V / 0.001 A.
- Calibrate current source: force through 100 Ω standard, measure V, actual I = V/100.
CHAPTER 5 — CALIBRATION PROCEDURE
- Zero the meter on a shorting bar: connect F+ to S+ and F− to S−. The meter should read <0.01 Ω. If not, apply offset null.
- Connect reference resistor in proper 4-wire configuration: F+ to one end, S+ to same end (inner), F− to other end, S− to same end (inner).
- Measure and record. Compare to certified value.
- Error ppm = (measured − certified) / certified × 106.
- Repeat for all decades: 0.1, 1, 10, 100, 1k Ω.
CHAPTER 6 — TUNING AND ADJUSTMENT
CHAPTER 7 — VERIFICATION
- Compare 10 Ω standard against Wheatstone bridge (TM-CAL-009).
- Agreement within 0.05% confirms both methods are consistent.
- Log all results; recertify standards annually.
APPENDIX A — CALCULATIONS AND FORMULAS
APPENDIX B — EXAMPLE RESULTS
| Rx nominal | 2-Wire (DMM) | 4-Wire (Kelvin) | Lead R (calculated) |
|---|---|---|---|
| 0.1 Ω | 0.42 Ω | 0.099 Ω | 0.16 Ω/lead |
| 1 Ω | 1.34 Ω | 1.001 Ω | 0.17 Ω/lead |
| 10 Ω | 10.34 Ω | 10.003 Ω | 0.17 Ω/lead |
| 100 Ω | 100.35 Ω | 100.02 Ω | 0.17 Ω/lead |