Unit 5 — Calibration and Verification

TM-GEAR-005 — Open Handout TM Chapters: Chapter 6, Chapter 7 ELOs: Calibrate the COAXIAL CABLE TRAPS — MULTIBAND ANTENNA ISOLATION; verify performance against acceptance criteria; complete calibration log Estimated time: 45 minutes (actual lab work required)


Step 1: Read the TM

Open TM-GEAR-005. Read Chapter 6 — Calibration and Chapter 7 — Verification and Acceptance completely.

Then come back here.


Chapter 6 — Calibration Content

  1. Measure trap Q: at resonance, measure the −3 dB bandwidth (BW) of the impedance peak. Q = f0 / BW. Target: Q ≥100.
  2. Verify isolation: connect the trap in series with a test transmission line. Measure S21 at f_trap. Isolation must be ≥20 dB at f_trap for effective band isolation.
  3. Measure insertion loss at off-resonant frequencies (other ham bands). Must be <0.5 dB at the bands the antenna is intended to operate on.

Chapter 7 — Verification Content

  1. Resonant frequency within ±100 kHz of target band center.
  2. Q ≥100 (BW ≤1% of center frequency).
  3. Isolation ≥20 dB at resonant frequency.
  4. Insertion loss <0.5 dB at all intended operating frequencies.
  5. Weather seal: immerse trap in water for 10 minutes; verify no water ingress to coax ends.
  6. Log: date, coax type, former OD, turns, measured f_trap, Q, isolation, operator.

Calibration vs. Verification — Distinction

Calibration (Chapter 6): active adjustments to bring the component within specification. You are permitted to adjust, tweak, and repeat until the criterion is met.

Verification (Chapter 7): independent confirmation that the calibration succeeded. No adjustments are permitted during verification. If a verification check fails, you return to Chapter 6, not to Chapter 7.


Acceptance Criterion

A criterion found in Chapter 6: 3 dB. Confirm the exact criterion in the TM.

Write the exact criterion from the TM here before you start the lab:



Lab Checklist

Work through each item in sequence. Do not skip steps.

  • [ ] All required equipment gathered and ready
  • [ ] Equipment warmed up per TM requirements
  • [ ] All Chapter 4 setup steps verified complete
  • [ ] Chapter 6 calibration performed (all steps)
  • [ ] Calibration criterion met (recorded below)
  • [ ] Chapter 7 verification performed (all steps, no adjustments)
  • [ ] All verification criteria passed
  • [ ] Calibration log entry completed

Calibration Log Entry

  • [ ] Date
  • [ ] Equipment used
  • [ ] Measurements taken
  • [ ] Results vs. criteria
  • [ ] Operator name

Record your results:

Item Measured Value Acceptance Criterion Pass/Fail
(from Ch 7)
(from Ch 7)
(from Ch 7)

Self-Check Questions

SC5-1. What is the acceptance criterion for this calibration (from Chapter 7)?

SC5-2. During verification, one reading fails. What is the correct action per the TM?

SC5-3. List the items required in a calibration log entry per Chapter 7.

SC5-4. What is the difference between calibration and verification in this context?

SC5-5. If the component passes all Chapter 7 checks, what does that tell you about its current state?


Answer Key

SC5-1. See Chapter 7. The criterion is stated as a specific value with units.

SC5-2. Stop verification. Return to Chapter 6 and re-calibrate. Diagnose the cause before re-adjusting.

SC5-3. See Chapter 7, last numbered step. The log items are listed there.

SC5-4. Calibration allows adjustments; verification does not. Verification is independent confirmation.

SC5-5. It tells you the component is performing within its specified parameters at this time. It does not tell you how long it will remain in calibration — periodic re-verification is required.


Checkpoint

You are ready for the final assessment when you can: - [ ] State the acceptance criterion from memory - [ ] State the difference between calibration and verification - [ ] State the required log items from memory

→ Proceed to Final Assessment