Unit 5 — Calibration and Verification
TM-GEAR-010 — Open Handout TM Chapters: Chapter 6, Chapter 7 ELOs: Calibrate the LIGHTNING PROTECTION AND STATION GROUNDING; verify performance against acceptance criteria; complete calibration log Estimated time: 45 minutes (actual lab work required)
Step 1: Read the TM
Open TM-GEAR-010. Read Chapter 6 — Calibration and Chapter 7 — Verification and Acceptance completely.
Then come back here.
Chapter 6 — Calibration Content
- Measure ground resistance: use a fall-of-potential test (three-point method) or a clamp-on ground resistance tester. Target: <10Ω per NEC; <5Ω recommended for RF stations.
- Verify continuity of all bonding straps with a low-resistance ohmmeter: any station ground conductor must read <0.1Ω end-to-end.
- GDT sparkover voltage: verify with a high-voltage tester if available. The sparkover voltage should match the manufacturer rating (typically 90–230 V depending on type).
Chapter 7 — Verification Content
- Ground resistance <5Ω (measure with fall-of-potential method).
- All bonding straps continuous (<0.1Ω).
- All coax feedlines pass through the entry panel with no direct path that bypasses the GDT.
- No RF current on coax shields within the shack (measure with clip-on RF current probe; target <10 mA during transmit on any feedline entering the shack).
- Log: date, ground rod resistance, number of GDT units installed and tested, bonding strap resistance, RF current check, operator.
Calibration vs. Verification — Distinction
Calibration (Chapter 6): active adjustments to bring the component within specification. You are permitted to adjust, tweak, and repeat until the criterion is met.
Verification (Chapter 7): independent confirmation that the calibration succeeded. No adjustments are permitted during verification. If a verification check fails, you return to Chapter 6, not to Chapter 7.
Acceptance Criterion
A criterion found in Chapter 6: 10Ω. Confirm the exact criterion in the TM.
Write the exact criterion from the TM here before you start the lab:
Lab Checklist
Work through each item in sequence. Do not skip steps.
- [ ] All required equipment gathered and ready
- [ ] Equipment warmed up per TM requirements
- [ ] All Chapter 4 setup steps verified complete
- [ ] Chapter 6 calibration performed (all steps)
- [ ] Calibration criterion met (recorded below)
- [ ] Chapter 7 verification performed (all steps, no adjustments)
- [ ] All verification criteria passed
- [ ] Calibration log entry completed
Calibration Log Entry
- [ ] Date
- [ ] Equipment used
- [ ] Measurements taken
- [ ] Results vs. criteria
- [ ] Operator name
Record your results:
| Item | Measured Value | Acceptance Criterion | Pass/Fail |
|---|---|---|---|
| (from Ch 7) | |||
| (from Ch 7) | |||
| (from Ch 7) |
Self-Check Questions
SC5-1. What is the acceptance criterion for this calibration (from Chapter 7)?
SC5-2. During verification, one reading fails. What is the correct action per the TM?
SC5-3. List the items required in a calibration log entry per Chapter 7.
SC5-4. What is the difference between calibration and verification in this context?
SC5-5. If the component passes all Chapter 7 checks, what does that tell you about its current state?
Answer Key
SC5-1. See Chapter 7. The criterion is stated as a specific value with units.
SC5-2. Stop verification. Return to Chapter 6 and re-calibrate. Diagnose the cause before re-adjusting.
SC5-3. See Chapter 7, last numbered step. The log items are listed there.
SC5-4. Calibration allows adjustments; verification does not. Verification is independent confirmation.
SC5-5. It tells you the component is performing within its specified parameters at this time. It does not tell you how long it will remain in calibration — periodic re-verification is required.
Checkpoint
You are ready for the final assessment when you can: - [ ] State the acceptance criterion from memory - [ ] State the difference between calibration and verification - [ ] State the required log items from memory
→ Proceed to Final Assessment