Unit 5 — Calibration and Verification

TM-GEAR-018 — Open Handout TM Chapters: Chapter 6, Chapter 7 ELOs: Calibrate the TRANSMATCH — PORTABLE ANTENNA TUNER; verify performance against acceptance criteria; complete calibration log Estimated time: 45 minutes (actual lab work required)


Step 1: Read the TM

Open TM-GEAR-018. Read Chapter 6 — Calibration and Chapter 7 — Verification and Acceptance completely.

Then come back here.


Chapter 6 — Calibration Content

  1. Verify SWR indicator zero: connect 50Ω load to the antenna port. Transmit 10W. Reflected LED must not illuminate. Adjust SWR bridge null trimmer if reflected shows any indication.
  2. Verify SWR indicator response: connect a 100Ω load (SWR 2:1). Reflected LED must illuminate at approximately 30% of the forward LED intensity (indicating ~2:1 SWR).

Chapter 7 — Verification Content

  1. Matches a 200Ω non-reactive load to SWR <1.5:1 on all HF bands.
  2. Matches a 450Ω random wire (via 9:1 un-un) to SWR <2:1 on all HF bands.
  3. Insertion loss with 50Ω load on both ports: <0.2 dB at 14 MHz.
  4. VHF section tunes 2M vertical to SWR <1.5:1 from a 75Ω feedpoint.
  5. All controls rotate smoothly over full range with no mechanical binding.
  6. Log: date, match test (200Ω at 7 and 14 MHz, SWR achieved), insertion loss check, VHF section test, operator.

Calibration vs. Verification — Distinction

Calibration (Chapter 6): active adjustments to bring the component within specification. You are permitted to adjust, tweak, and repeat until the criterion is met.

Verification (Chapter 7): independent confirmation that the calibration succeeded. No adjustments are permitted during verification. If a verification check fails, you return to Chapter 6, not to Chapter 7.


Acceptance Criterion

A criterion found in Chapter 6: 30%. Confirm the exact criterion in the TM.

Write the exact criterion from the TM here before you start the lab:



Lab Checklist

Work through each item in sequence. Do not skip steps.

  • [ ] All required equipment gathered and ready
  • [ ] Equipment warmed up per TM requirements
  • [ ] All Chapter 4 setup steps verified complete
  • [ ] Chapter 6 calibration performed (all steps)
  • [ ] Calibration criterion met (recorded below)
  • [ ] Chapter 7 verification performed (all steps, no adjustments)
  • [ ] All verification criteria passed
  • [ ] Calibration log entry completed

Calibration Log Entry

  • [ ] Date
  • [ ] Equipment used
  • [ ] Measurements taken
  • [ ] Results vs. criteria
  • [ ] Operator name

Record your results:

Item Measured Value Acceptance Criterion Pass/Fail
(from Ch 7)
(from Ch 7)
(from Ch 7)

Self-Check Questions

SC5-1. What is the acceptance criterion for this calibration (from Chapter 7)?

SC5-2. During verification, one reading fails. What is the correct action per the TM?

SC5-3. List the items required in a calibration log entry per Chapter 7.

SC5-4. What is the difference between calibration and verification in this context?

SC5-5. If the component passes all Chapter 7 checks, what does that tell you about its current state?


Answer Key

SC5-1. See Chapter 7. The criterion is stated as a specific value with units.

SC5-2. Stop verification. Return to Chapter 6 and re-calibrate. Diagnose the cause before re-adjusting.

SC5-3. See Chapter 7, last numbered step. The log items are listed there.

SC5-4. Calibration allows adjustments; verification does not. Verification is independent confirmation.

SC5-5. It tells you the component is performing within its specified parameters at this time. It does not tell you how long it will remain in calibration — periodic re-verification is required.


Checkpoint

You are ready for the final assessment when you can: - [ ] State the acceptance criterion from memory - [ ] State the difference between calibration and verification - [ ] State the required log items from memory

→ Proceed to Final Assessment