Unit 5 — Calibration and Verification
TM-TOOL-002 — Open Handout TM Chapters: Chapter 6, Chapter 7 ELOs: Calibrate the ANTENNA ANALYZER — VECTOR IMPEDANCE MEASUREMENT; verify performance against acceptance criteria; complete calibration log Estimated time: 45 minutes (actual lab work required)
Step 1: Read the TM
Open TM-TOOL-002. Read Chapter 6 — Calibration and Chapter 7 — Verification and Acceptance completely.
Then come back here.
Chapter 6 — Calibration Content
6-1 SOLT Calibration
Before measuring an antenna, perform a one-port SOLT calibration at the DUT SMA connector (not at the instrument chassis):
- Connect SHORT (shorted SMA cap). Press CAL → SHORT.
- Connect OPEN (SMA cap with no connection). Press CAL → OPEN.
- Connect 50Ω LOAD (SMA terminator). Press CAL → LOAD.
- Press CAL → DONE. Calibration plane is now at DUT connector.
After SOLT cal, short should read SWR >50:1, open should read SWR >50:1, and 50Ω load should read SWR <1.05:1 across the calibrated band.
Chapter 7 — Verification Content
- Connect a precision 50Ω load (Pasternack PE6010 or calibrated terminator). Verify SWR <1.05:1 at all calibrated frequencies.
- Connect a known 100Ω resistor (1%, non-inductive). Verify R reads 100 ± 5Ω, X reads 0 ± 5Ω at 1 MHz.
- Connect a 50Ω + 50 nH load (resistor in series with known inductor). Verify jX reads within ±10% of calculated inductive reactance at 10 MHz.
- If any check fails, repeat SOLT calibration. Persistent errors indicate coupler asymmetry or AD8302 bias fault.
- Log: date, calibration kit used, frequency range, short SWR, open SWR, load SWR, 100Ω R/X readings.
Calibration vs. Verification — Distinction
Calibration (Chapter 6): active adjustments to bring the tool within specification. You are permitted to adjust, tweak, and repeat until the criterion is met.
Verification (Chapter 7): independent confirmation that the calibration succeeded. No adjustments are permitted during verification. If a verification check fails, you return to Chapter 6, not to Chapter 7.
Acceptance Criterion
A criterion found in Chapter 6: 50Ω. Confirm the exact criterion in the TM.
Write the exact criterion from the TM here before you start the lab:
Lab Checklist
Work through each item in sequence. Do not skip steps.
- [ ] All required equipment gathered and ready
- [ ] Equipment warmed up per TM requirements
- [ ] All Chapter 4 setup steps verified complete
- [ ] Chapter 6 calibration performed (all steps)
- [ ] Calibration criterion met (recorded below)
- [ ] Chapter 7 verification performed (all steps, no adjustments)
- [ ] All verification criteria passed
- [ ] Calibration log entry completed
Calibration Log Entry
- [ ] Date
- [ ] Equipment used
- [ ] Measurements taken
- [ ] Results vs. criteria
- [ ] Operator name
Record your results:
| Item | Measured Value | Acceptance Criterion | Pass/Fail |
|---|---|---|---|
| (from Ch 7) | |||
| (from Ch 7) | |||
| (from Ch 7) |
Self-Check Questions
SC5-1. What is the acceptance criterion for this calibration (from Chapter 7)?
SC5-2. During verification, one reading fails. What is the correct action per the TM?
SC5-3. List the items required in a calibration log entry per Chapter 7.
SC5-4. What is the difference between calibration and verification in this context?
SC5-5. If the tool passes all Chapter 7 checks, what does that tell you about its current state?
Answer Key
SC5-1. See Chapter 7. The criterion is stated as a specific value with units.
SC5-2. Stop verification. Return to Chapter 6 and re-calibrate. Diagnose the cause before re-adjusting.
SC5-3. See Chapter 7, last numbered step. The log items are listed there.
SC5-4. Calibration allows adjustments; verification does not. Verification is independent confirmation.
SC5-5. It tells you the tool is performing within its specified accuracy at this time. It does not tell you how long it will remain in calibration — periodic re-calibration is required.
Checkpoint
You are ready for the final assessment when you can: - [ ] State the acceptance criterion from memory - [ ] State the difference between calibration and verification - [ ] State the required log items from memory
→ Proceed to Final Assessment