Unit 5 — Calibration and Verification
TM-TOOL-007 — Open Handout TM Chapters: Chapter 6, Chapter 7 ELOs: Calibrate the COMPACT PORTABLE RF POWER METER; verify performance against acceptance criteria; complete calibration log Estimated time: 45 minutes (actual lab work required)
Step 1: Read the TM
Open TM-TOOL-007. Read Chapter 6 — Calibration and Chapter 7 — Verification and Acceptance completely.
Then come back here.
Chapter 6 — Calibration Content
- Connect a calibrated RF power source (TinySA in generator mode, or signal generator with known output) to the IN port. Load OUT with a calibrated 50Ω dummy load.
- Apply −10 dBm, 0 dBm, +10 dBm, +20 dBm at 14 MHz (HF calibration). Record ADC voltage vs. known power.
- Fit a calibration curve (linear or polynomial) to the V_det vs. P_in data. Store coefficients in ESP32 NVS. Accuracy after calibration: ±5% over the calibrated range.
- Repeat calibration procedure at 144 MHz for the VHF/UHF section.
Chapter 7 — Verification Content
- Apply known 5W at 14.175 MHz. Display must show 5.0 ± 0.3W.
- Apply known 5W at 145.0 MHz. Display must show 5.0 ± 0.3W.
- Short OUT connector (SWR = ∞). Verify CYD shows SWR ≥99:1 and forward/reflected power are approximately equal.
- Insertion loss verification: compare power reading with coupler in-line vs. direct connection. Difference must be <0.2 dB.
- Log: date, calibration source, power levels checked, measured vs. expected values, operator.
Calibration vs. Verification — Distinction
Calibration (Chapter 6): active adjustments to bring the tool within specification. You are permitted to adjust, tweak, and repeat until the criterion is met.
Verification (Chapter 7): independent confirmation that the calibration succeeded. No adjustments are permitted during verification. If a verification check fails, you return to Chapter 6, not to Chapter 7.
Acceptance Criterion
A criterion found in Chapter 6: 5%. Confirm the exact criterion in the TM.
Write the exact criterion from the TM here before you start the lab:
Lab Checklist
Work through each item in sequence. Do not skip steps.
- [ ] All required equipment gathered and ready
- [ ] Equipment warmed up per TM requirements
- [ ] All Chapter 4 setup steps verified complete
- [ ] Chapter 6 calibration performed (all steps)
- [ ] Calibration criterion met (recorded below)
- [ ] Chapter 7 verification performed (all steps, no adjustments)
- [ ] All verification criteria passed
- [ ] Calibration log entry completed
Calibration Log Entry
- [ ] Date
- [ ] Equipment used
- [ ] Measurements taken
- [ ] Results vs. criteria
- [ ] Operator name
Record your results:
| Item | Measured Value | Acceptance Criterion | Pass/Fail |
|---|---|---|---|
| (from Ch 7) | |||
| (from Ch 7) | |||
| (from Ch 7) |
Self-Check Questions
SC5-1. What is the acceptance criterion for this calibration (from Chapter 7)?
SC5-2. During verification, one reading fails. What is the correct action per the TM?
SC5-3. List the items required in a calibration log entry per Chapter 7.
SC5-4. What is the difference between calibration and verification in this context?
SC5-5. If the tool passes all Chapter 7 checks, what does that tell you about its current state?
Answer Key
SC5-1. See Chapter 7. The criterion is stated as a specific value with units.
SC5-2. Stop verification. Return to Chapter 6 and re-calibrate. Diagnose the cause before re-adjusting.
SC5-3. See Chapter 7, last numbered step. The log items are listed there.
SC5-4. Calibration allows adjustments; verification does not. Verification is independent confirmation.
SC5-5. It tells you the tool is performing within its specified accuracy at this time. It does not tell you how long it will remain in calibration — periodic re-calibration is required.
Checkpoint
You are ready for the final assessment when you can: - [ ] State the acceptance criterion from memory - [ ] State the difference between calibration and verification - [ ] State the required log items from memory
→ Proceed to Final Assessment