Unit 4 — Calibration and Tuning
TM-ANT-058 — Open Handout TM Chapters: Chapter 5, Chapter 6, Appendix A ELOs: Calibrate the SLINKY ANTENNA — 20M PORTABLE to resonance; tune SWR to within acceptance criteria; apply frequency calculations Estimated time: 45 minutes (actual lab work required)
Step 1: Read the TM
Open TM-ANT-058. Read Chapter 5 — Calibration Procedure, Chapter 6 — Tuning and Adjustment, and Appendix A completely.
Then come back here.
Chapter 5 — Calibration Procedure
5-1. NANOVNA SWR AND IMPEDANCE SWEEP
NOTE: The NEC2 model file for this antenna is included in the antenna directory. Run it with xnec2c, 4nec2, or any NEC2-compatible engine to generate polar plots, impedance data, and gain figures. The NanoVNA measurements in Chapter 5 should be compared against NEC2 predictions — deviations >3 dB or >20% impedance indicate a construction error.
- Perform SOLT calibration on NanoVNA using the SOL (Short-Open-Load) kit at the antenna end of the feed line.
- Set NanoVNA sweep range to cover ±10% of target center frequency (example: 40M → 6.5–7.8 MHz).
- Connect NanoVNA to feedpoint. Navigate to CH0 S11 display. Select Smith Chart and SWR graphs.
- Record: frequency of minimum SWR (fres), SWR at fres, SWR at band edges, R + jX at fres.
- Resonance is confirmed when X ≈ 0 and R ≈ 73 Ω (simple dipole) or per design (matched system).
- Compare measured fres to design frequency. If fres is too high, the element is short — lengthen each arm 1–2 in. If fres is too low, the element is long — trim each arm 1 in. Repeat until fres is within ±0.5% of design frequency.
Chapter 6 — Tuning and Adjustment
6-1. RESONANCE ADJUSTMENT
Trim or extend element arms symmetrically to shift resonance. Each 1-inch change in total length shifts resonance by approximately fMHz/468 × 12 kHz for a 40M dipole. Adjust in 2-inch increments and re-measure SWR between adjustments. Finalize element length when SWR at design frequency is <1.5:1 or as specified.
6-2. IMPEDANCE MATCHING
If impedance at resonance differs from 50 Ω, adjust element height (increases ground effect), add a matching network (L-network, λ/4 transformer, or series capacitor), or use a 4:1 balun for designs with higher feed-point impedance such as folded dipoles (50 Ω).
Appendix A — Calculations and Formulas
Half-wave dipole element length (each arm, feet)
Larm = 234 / fMHz
Velocity factor correction (for insulated wire)
Ladj = Larm × Vf (Vf ≈ 0.95 for #14 AWG PVC)
Radiation resistance (half-wave dipole, free space)
Rr = 73.1 Ω
Gain over isotropic (free space)
G = 2.14 dBi
Height for DX angle (degrees elevation for max gain)
θ ≈ arcsin(1 / (2h/λ)) for h > 0.25λ
Key Formulas Summary
Larm = 234 / fMHzLadj = Larm × Vf (Vf ≈ 0.95 for #14 AWG PVC)Rr = 73.1 ΩG = 2.14 dBi
Calibration vs. Tuning — Distinction
Calibration (Chapter 5): initial setup measurements to establish current antenna state — resonant frequency, SWR at design frequency, impedance at feedpoint. No adjustments yet; you are characterizing what you built.
Tuning (Chapter 6): active adjustments based on calibration data. Trim the element, adjust the matching network, or change height/orientation until Chapter 7 acceptance criteria are met.
Acceptance Criterion
Chapter 5 specifies a criterion: 3 dB. Confirm the exact value in the TM.
Write the exact criterion from the TM here before you start:
Lab Checklist
- [ ] Antenna assembled per Chapter 4 and installed at operating height
- [ ] NanoVNA or SWR bridge connected per Chapter 5
- [ ] Chapter 5 calibration measurements taken and recorded
- [ ] Resonant frequency identified (minimum SWR point)
- [ ] SWR at design frequency recorded
- [ ] Chapter 6 tuning performed if needed
- [ ] Acceptance criterion met
Calibration Log
| Parameter | Measured | Specification | Pass/Fail |
|---|---|---|---|
| Resonant frequency (MHz) | (design freq ± tolerance) | ||
| SWR at design frequency | (from Ch. 5) | ||
| Feedpoint impedance (Ω) | (from Ch. 2) |
Practice Problems
P4-1. The design frequency of the SLINKY ANTENNA — 20M PORTABLE is stated in Chapter 1. Using the formula from Appendix A, compute the theoretical element length for that frequency. Show all work.
P4-2. Your NanoVNA shows minimum SWR at a frequency 3% above the design frequency. Which direction do you adjust — lengthen or shorten? By what percentage?
P4-3. SWR at resonance is 1.8:1, but the acceptance criterion is SWR ≤ 2.0:1. Does the antenna pass? What does SWR 1.8:1 mean in terms of reflected power?
P4-4. Compute reflected power percentage for SWR = 2.0:1. Formula: reflected power (%) = ((SWR − 1)/(SWR + 1))² × 100
Answer Key — Practice Problems
P4-1. See Appendix A. For a dipole: L(ft) = 468/f(MHz). For a quarter-wave vertical: L(ft) = 234/f(MHz). Apply the formula specific to this antenna type.
P4-2. Resonance is above design frequency → antenna is electrically short → lengthen the element. Adjustment magnitude: approx. 3% longer.
P4-3. SWR 1.8:1 passes the 2.0:1 criterion. Reflected power: ((1.8−1)/(1.8+1))² × 100 = (0.8/2.8)² × 100 = 8.2% reflected.
P4-4. SWR 2.0:1: ((2−1)/(2+1))² × 100 = (1/3)² × 100 = 11.1% reflected. Most of the power still gets through.
Checkpoint
Before proceeding: - [ ] Calibration measurements taken and logged - [ ] Antenna tuned to within the acceptance criterion - [ ] You can compute element length adjustment from frequency error
→ Proceed to Unit 5