Unit 4 — Calibration Procedure and Calculations
TM-CAL-015 — Open Handout TM Chapters: Chapter 5, Appendix A ELOs: Execute calibration procedure; compute error and determine pass/fail Estimated time: 30 minutes
Step 1: Read the TM
Open TM-CAL-015. Read Chapter 5 and Appendix A completely.
Then come back here.
Chapter 5 — Calibration Procedure
- Connect pad to TinySA: source on input, TinySA on output.
- Set TinySA to 0–200 MHz, measure insertion loss.
- Verify attenuation = Ameasured at multiple frequencies: 1, 10, 30, 100, 200 MHz.
- Attenuation should be flat within ±0.2 dB from DC to 100 MHz.
- Measure input reflection: |S11| should be <−20 dB (SWR <1.22) if resistor values are correct.
- Record actual attenuation vs. designed value. Correction = Adesigned − Ameasured.
Appendix A — Formulas
Attenuation voltage ratio
k = 10AdB/20
Pi shunt resistors
Rshunt = Z0(k+1)/(k−1)
Pi series resistor
Rseries = Z0 × 2k / (k2−1)
Total power at input for measuring attenuation (dB from V ratio)
AdB = 20 × log10(Vin/Vout)
Key Formulas Summary
- k = 10AdB/20
- Rshunt = Z0(k+1)/(k−1)
- Rseries = Z0 × 2k / (k2−1)
- AdB = 20 × log10(Vin/Vout)
The Calibration Procedure
Chapter 5 specifies 6 calibration steps.
Calibration is a comparison: you apply a known reference value to the instrument under test and record what the instrument reads. The difference is the error. You then either: - Adjust the instrument until the error is within the acceptance criterion, or - Record the error as a correction factor to apply to future readings
Locate the acceptance criterion in Chapter 5 or Chapter 7.
Practice Problems
Work these before checking answers.
P4-1. The reference value is 10.000 V. Your instrument reads 10.043 V. (a) What is the error in volts? (b) What is the error as a percentage of the reference value? Show your work.
P4-2. Using the formula for % error: error% = (measured − reference) / reference × 100 Apply it to: reference = 100.0 kHz, measured = 99,985 Hz. (a) Error in Hz. (b) Error in %. (c) Error in ppm.
P4-3. The acceptance criterion in the TM is ±1%. Your measurement gives an error of +0.8%. Does it pass? State your reasoning.
Practice Problem Answers
P4-1. (a) 10.043 − 10.000 = +0.043 V (b) 0.043 / 10.000 × 100 = +0.43%
P4-2. (a) 99,985 − 100,000 = −15 Hz (b) −15 / 100,000 × 100 = −0.015% (c) −15 / 100,000 × 1,000,000 = −150 ppm
P4-3. +0.8% is within ±1%. PASS. State: "error = +0.8%; criterion = ±1%; result = PASS." Always cite the TM section for the criterion.
Self-Check Questions
SC4-1. How many steps does Chapter 5 specify for the calibration procedure?
SC4-2. What reference value(s) does Chapter 5 apply to the instrument under test?
SC4-3. State the calibration acceptance criterion from the TM. Cite the section.
SC4-4. Write the error formula from Appendix A. Include units.
SC4-5. If the instrument reads 2.3% high, is the error positive or negative? What does a positive error indicate?
Answer Key
SC4-1. Count the numbered steps in Chapter 5. (TM Ch. 5)
SC4-2. See Chapter 5 for the reference values applied. These are the known-good inputs used to check the instrument. (TM §5-1)
SC4-3. See TM Chapter 5 or Chapter 7. Copy the criterion exactly with units and section number.
SC4-4. See Appendix A. Write it exactly as shown.
SC4-5. Positive (instrument reads higher than reference). A positive error means the instrument over-reads — it reports a value higher than the true value. (TM App. A)
Checkpoint
Before proceeding, you must be able to: - State the calibration acceptance criterion without looking - Write the error formula from memory - Work a % error and ppm calculation correctly
→ Proceed to Unit 5