Unit 4 — Calibration Procedure and Calculations

TM-CAL-013 — Open Handout TM Chapters: Chapter 5, Appendix A ELOs: Execute calibration procedure; compute error and determine pass/fail Estimated time: 30 minutes


Step 1: Read the TM

Open TM-CAL-013. Read Chapter 5 and Appendix A completely.

Then come back here.


Chapter 5 — Calibration Procedure

  1. Calculate expected capacitance: C = (pF/ft) × length(ft).
  2. Example: 1 m of RG-58 = 3.281 ft × 28.5 pF/ft = 93.5 pF.
  3. Measure with DMM capacitance function or LCR meter: connect shorting BNC to one end, measure at open BNC.
  4. Compare measured vs. calculated. Error reflects meter accuracy.
  5. Error = (measured − calculated) / calculated × 100%.
  6. If meter reads 91.2 pF and calculated is 93.5 pF: error = −2.5%.
  7. Apply correction factor to future capacitance measurements.

Appendix A — Formulas

Total capacitance
Ctotal = (C/m) × lmeters = (C/ft) × lfeet
RG-58, 1.524 m (5 ft)
C = 28.5 × 5 = 142.5 pF
Velocity factor and capacitance relationship
VF = 1/√εr, so C/m = 1/(Z0 × VF × c)

Key Formulas Summary

  • Ctotal = (C/m) × lmeters = (C/ft) × lfeet
  • C = 28.5 × 5 = 142.5 pF
  • VF = 1/√εr, so C/m = 1/(Z0 × VF × c)

The Calibration Procedure

Chapter 5 specifies 7 calibration steps.

Calibration is a comparison: you apply a known reference value to the instrument under test and record what the instrument reads. The difference is the error. You then either: - Adjust the instrument until the error is within the acceptance criterion, or - Record the error as a correction factor to apply to future readings

An error criterion found in Chapter 5: 100%. Confirm the exact criterion in the TM.


Practice Problems

Work these before checking answers.

P4-1. The reference value is 10.000 V. Your instrument reads 10.043 V. (a) What is the error in volts? (b) What is the error as a percentage of the reference value? Show your work.

P4-2. Using the formula for % error: error% = (measured − reference) / reference × 100 Apply it to: reference = 100.0 kHz, measured = 99,985 Hz. (a) Error in Hz. (b) Error in %. (c) Error in ppm.

P4-3. The acceptance criterion in the TM is ±1%. Your measurement gives an error of +0.8%. Does it pass? State your reasoning.


Practice Problem Answers

P4-1. (a) 10.043 − 10.000 = +0.043 V (b) 0.043 / 10.000 × 100 = +0.43%

P4-2. (a) 99,985 − 100,000 = −15 Hz (b) −15 / 100,000 × 100 = −0.015% (c) −15 / 100,000 × 1,000,000 = −150 ppm

P4-3. +0.8% is within ±1%. PASS. State: "error = +0.8%; criterion = ±1%; result = PASS." Always cite the TM section for the criterion.


Self-Check Questions

SC4-1. How many steps does Chapter 5 specify for the calibration procedure?

SC4-2. What reference value(s) does Chapter 5 apply to the instrument under test?

SC4-3. State the calibration acceptance criterion from the TM. Cite the section.

SC4-4. Write the error formula from Appendix A. Include units.

SC4-5. If the instrument reads 2.3% high, is the error positive or negative? What does a positive error indicate?


Answer Key

SC4-1. Count the numbered steps in Chapter 5. (TM Ch. 5)

SC4-2. See Chapter 5 for the reference values applied. These are the known-good inputs used to check the instrument. (TM §5-1)

SC4-3. See TM Chapter 5 or Chapter 7. Copy the criterion exactly with units and section number.

SC4-4. See Appendix A. Write it exactly as shown.

SC4-5. Positive (instrument reads higher than reference). A positive error means the instrument over-reads — it reports a value higher than the true value. (TM App. A)


Checkpoint

Before proceeding, you must be able to: - State the calibration acceptance criterion without looking - Write the error formula from memory - Work a % error and ppm calculation correctly

→ Proceed to Unit 5