Unit 2 — Equipment and Test Setup

TM-CAL-002 — Open Handout TM Chapter: Chapter 3 ELOs: Identify required equipment; understand test configuration Estimated time: 20 minutes


Step 1: Read the TM

Open TM-CAL-002. Read Chapter 3 — Materials and Construction completely.

Then come back here.


Chapter 3 Content

3-1. BILL OF MATERIALS

Qty Item Specification Notes
1 GPS receiver module NEO-M8N preferred; NEO-6M acceptable Must have 1PPS output pin
1 GPS patch antenna Active, 25×25 mm or 35×35 mm Magnetic mount for mobility
1 SMA or IPEX cable 50 Ω, length as needed GPS antenna to module
1 5 V power supply 100 mA minimum, clean DC USB supply acceptable
1 SMA/BNC adapter For 1PPS output Or BNC connector direct
1 50 Ω BNC cable RG-58, 0.5–1 m 1PPS to counter/scope

3-2. LEVEL SHIFTING

If the GPS module outputs 3.3 V logic and the counter requires 5 V TTL, a simple level-shift circuit is needed.

Voltage divider for 5V to 3.3V (GPS antenna power only)
Not required for 1PPS: most counters accept 3.3 V on TTL input

NOTE: The NanoVNA, TinySA Ultra, and most modern frequency counters accept 3.3 V logic on TTL inputs. No level shifting is required for these instruments.


Equipment Readiness

Review the equipment list in Chapter 3 carefully.

Before beginning any calibration, verify every item on the equipment list is: - Present and in working order - Within its own calibration period (do not calibrate with an uncalibrated reference) - Set up per the configuration specified in Chapter 3

The reference quality is the ceiling on your calibration quality. A reference with unknown accuracy produces a calibration with unknown accuracy.


Self-Check Questions

SC2-1. List the three most important pieces of equipment specified in Chapter 3.

SC2-2. What reference standard(s) does the TM specify for this calibration?

SC2-3. Does Chapter 3 specify any warm-up time or settling requirement? If so, state it.

SC2-4. What is the test configuration? (How is the equipment connected or arranged?)

SC2-5. What would be the consequence of using a reference standard that is itself out of calibration?


Answer Key

SC2-1. See Chapter 3 equipment table or list. Identify the three items that appear first or are marked most critical.

SC2-2. See Chapter 3. The reference standard is the source of known-good values used to check the instrument under test.

SC2-3. See Chapter 3. Warm-up time is often specified for instruments that use oscillators or amplifiers — they need thermal stabilization before measurements are valid.

SC2-4. See Chapter 3. Describe the connection or arrangement in your own words, then verify against the TM.

SC2-5. The calibration error of the reference propagates directly into your calibration result. If the reference is 5% off and you do not know it, your calibration will also be 5% off — and you will not know that either.


Checkpoint

Before proceeding, you must be able to: - List the required equipment from memory - State the reference standard and its specified accuracy

→ Proceed to Unit 3