Self-Paced Course — LC RESONANCE STANDARD

TM-CAL-012 — Open Handout TM-CAL-012 Estimated time: 2–3 hours (reading + practice + lab) Handout: TM-CAL-012 (lc_resonance_standard/tm.html) — keep it open throughout


What You Will Be Able to Do

TLO-1: Given the required equipment — deploy and use the LC RESONANCE STANDARD to verify or calibrate test instruments per TM-CAL-012, achieving the traceability and accuracy stated in Chapter 7, with a completed calibration log entry.

TLO-2: Given measured and reference values — compute the calibration error in Hz and ppm (or appropriate units) and determine pass/fail against the stated criterion.


Scope

This manual covers construction of LC resonant circuits as fixed frequency standards and markers. When constructed with stable components (toroid inductors, NP0 capacitors), resonant frequency accurac


What You Need

Handout

  • TM-CAL-012 (LC RESONANCE STANDARD) — open alongside this course

Equipment (needed for Unit 5 lab)

  • Reference standards and instruments specified in Chapter 3
  • Calculator

Course Units

Unit Topic TM Chapters Est. Time
Unit 1 Theory of Operation Ch. 2 20 min
Unit 2 Equipment and Test Setup Ch. 3 20 min
Unit 3 Setup Procedures Ch. 4 15 min
Unit 4 Calibration Procedure and Calculations Ch. 5, App. A 30 min
Unit 5 Verification and Logging Ch. 6, 7 45 min (lab)
Final Assessment 15-question written test All 20 min

How to Use This Course

  1. Work units in order.
  2. For each unit: read the assigned TM chapter first, then read the unit, then answer the self-check questions.
  3. Self-check answers are at the bottom of each unit — try before looking.
  4. The checkpoint at the end of each unit is a go/no-go.
  5. Complete the Unit 5 lab before taking the final assessment.
  6. Passing score: 13/15 (87%). All calculation questions must be correct regardless of total score.